Effect of temperature on calibration quality of structured-light three-dimensional scanners

Paper 01

Effect of temperature on calibration quality of structured-light three-dimensional scanners

Abstract: This paper presents the outcome of research into the effects of ambient temperature changes on structured-light three-dimensional (3D) scanners. The tests were conducted in a thermal chamber and consisted of a comparison of the 3D measurement of a special reference unit (made of a carbon composite) performed at different temperatures, with measurements performed at the calibration temperature. A contact measuring arm with temperature compensation was used as a reference. Based on the results of these experiments, we propose a method that allows us to extend the existing scanner calibration method by using a temperature-correction procedure that is based on linear and nonlinear mathematical models. An exemplary application of this procedure has shown that the range of temperatures in which scanner accuracy is within declared limits can be increased 11-fold.

Journal: Applied Optics

Authors: Marcin Adamczyk, Michał Kamiński, Robert Sitnik, Anna Bogdan, Maciej Karaszewski

Year: 2014

DOI: 10.1364/AO.53.005154

 

 